Loading…

Variable Name: device_specific_status_enum_0

Variable ID: 0x4000

Variable Label: Error status

Variable Help:

Variable Handling: R

Variable Class: Dynamic | Diagnostic | HART

Variable Size: 1

Variable DataType: Enum

Variable Enum Value Variable Enum Text
0x0000 OK
0x0001 Ultrasonic signal too large. Amplifier overload.
0x0002 Ultrasonic signal too small. Check ultrasonic sensor connection.
0x0003 Wall thickness exceeds measurement range.
0x0004 Thickness under range. Minimum thickness – 3mm.
0x0005 Battery voltage low. Replace battery.
0x0006 Averaging failure.
0x0007 Memory not cleared. Power control failure.
0x0008 Parameter outside expected range.
0x0009 Sensor Board Failed To Respond.
0x000A Back Wall Echo too small. Improve sensor bonding to pipe.
0x000B Front Wall Echo too small or late arrival. Retry measurement. Potential sensor problem

Variable Display Format:

Variable Edit Format:

Variable Unit:

Variable Max: N/A

Variable Min: N/A

Variable Read Command List

Command Number Command ID Command Name
48 1043 read_additional_device_status

Variable Write Command List

None