Variable Name: device_specific_status_enum_0
Variable ID: 0x4000
Variable Label: Error status
Variable Help:
Variable Handling: R
Variable Class: Dynamic | Diagnostic | HART
Variable Size: 1
Variable DataType: Enum
Variable Enum Value | Variable Enum Text |
---|---|
0x0000 | OK |
0x0001 | Ultrasonic signal too large. Amplifier overload. |
0x0002 | Ultrasonic signal too small. Check ultrasonic sensor connection. |
0x0003 | Wall thickness exceeds measurement range. |
0x0004 | Thickness under range. Minimum thickness 3mm. |
0x0005 | Battery voltage low. Replace battery. |
0x0006 | Averaging failure. |
0x0007 | Memory not cleared. Power control failure. |
0x0008 | Parameter outside expected range. |
0x0009 | Sensor Board Failed To Respond. |
0x000A | Back Wall Echo too small. Improve sensor bonding to pipe. |
0x000B | Front Wall Echo too small or late arrival. Retry measurement. Potential sensor problem |
Variable Display Format:
Variable Edit Format:
Variable Unit:
Variable Max: N/A
Variable Min: N/A
Variable Read Command List
Command Number | Command ID | Command Name |
---|---|---|
48 | 1043 | read_additional_device_status |
Variable Write Command List
None